|
Thermo-Conversion Elemental Analyzer (TCEA) Standard Operating ProceduresPhase 1: Perform TCEA Maintenance1. Replace the graphite crucible every 75-100 samples. Phase 2: Warm-Up the day prior to running1. Close SGE valve at MS source & turn on Conflo He dilution.2. Make sure SGE valve for gas bench, GC, or LC valve is also closed. 3. Clean glassy carbon tube and glassy carbon grains if needed (see below). 4. Set the temperature controller ramp rate.
6. Set Jumbo controller to ramp up to the desired operating temperature overnight.
Phase 2: The day of the run0. If running phosphates, procure dry ice (located in 137 Sinsheimer and in the hallway across from 339 PSB, delivered every Tuesday and Friday).1. Set He and Purge He carrier gas flow to 1.7 bar. 2. Open the CO tank and regulator valves in the fume hood. Check both the cylinder valve and the regulator valve are open. (Temporary: attach the Variac power supply to the automatic shutoff valve on the CO line in the fume hood. This is to bypass a defective circuit powering this valve). 3. Open CO gas tanks by opening the valves below the peripheals benchtop. 4. Close the isolation valve on the zero blank autosamper (ZBA). 5. Load samples and vent the ZBA carousel cavity
8. Set the communication switch box: Select Sherlock/Columbo EA/TCEA communication switch box according to the instrument being used. 9. On the Conflo, turn the dilution on. 10. Turn off source at the computer and then open the SGE valve at the mass spectrometer source. Be sure the other peripheral SGE valves are closed. 11. Check that the vacuum pressure is between 7-e7 to 3e-6 mBar range, and then turn the source back on. If outside this range, close the source SGE valve and see the lab manager. 12. Select CO gas on the mass spectrometer software, and monitor mass 28 background. If it is less than 1000 mV, then turn off dilution. 13. Open the Instrument Control software module and scan mass 28 for background stability. Wait until the background trace flattens out and stabilizes, then stop scan. 14. Return to Acquisition software and begin a standard on-off sequence for CO. You should achieve better than 0.1‰ precision for the standard on off δ18O. 15. While the standard on-off are going, prepare a sequence file for the run. 16. If running phosphates, crush up some dry ice and add it to ethanol, to make a dry ice-ethanol slush. Add this to slush to a small glass dewar (found at the back of the TCEA peripheral). Place the coiled 1/16 stainless steel line into the trap, such that the coil is at the bottom of the dewar. 17. Turn the 2-way valve, that is downstream of the coil at the back of the TCEA peripheral, such that gas from the TECA travels towards the mass spectrometer and not to waste. 18. Start sequence. 19. Log use on the SIL web-based form. 20. Monitor at least the first few standards to ensure instrument is performing properly. Run Shutdown:1. Close SGE valve at the mass spectrometer source.2. Close CO valve on front of the peripheral benchtop. Close CO valves at the gas cylinder and regulator located in the fume hood. Shut off the Variac too (if necessary). 3. Turn on the Conflo III dilution (Isodat software). 4. Set the temperature controller ramp rate to 3°C/min:
Note: Be very careful that the pressure doesnt drop below this flow rate. The carrier flow drops very slowly and operators have allowed the flow to drop to zero, at which point the furnace control is shut-off and resulting in rapid cooling of the furnace. Once the pressure reaches 0.6 bar, be sure that if you can increase the pressure by turning up the regulator, to confirm the pressure is not still dropping. Rapid uncontrolled cooling can cause very expensive repair and instrument down-time. 6. Set Jumbo controller to ramp down to room temperature:
top of page Maintenance Procedures:1. Replace the graphite crucible every 75-100 samples. A crucible can be removed if the temperature of the furnace is in standby conditions (i.e., with the furnace temperature at 600°C), but not at operating temperature!2. Remove the reactor and clean the glassy carbon tube and glassy carbon grains after every 200-250 samples.
5. Reassemble the reactor column.
Note: Failure to achieve a leak tight connection at this point likely means you will need to cool the furnace down the following day, and will loose 1-2 days without any analyses. |
![]() NEWS & EVENTSUCSC's SIL retires its 15 year-old dual-inlet Optima following a very productive career that generated over 40,000 data points. Two Tsavo Lions In Famed Killings Get Partial Reprieve according to UCSC Graduate Student Justin Yeakel as inteviewed on NPR's All Things Considered. Listen here: SIL co-director Christina Ravelo (Ocean Sciences) sails as co-chief scientist on Integrated Ocean Drilling Project (IODP) Expedition 323 to investigate Bearing Sea Paleoceanography.
SIL co-director Christina Ravelo (Ocean Sciences) gives the 2008 Emiliani Lecture at the American Geophysical Union Meeting in San Francisco on "Lessons from the Pliocene Warm Period and the Onset of Northern Hemisphere Glaciation".
UCSC SIL has been funded by the National Science Foundation for a new Dual-Inlet Isotope Ratio Mass Spectrometer with individual acid drip system for very small calcium carbonate samples. SIL co-director Jim Zachos (Earth and Planetary Sciences) recieves prestigious Humbolt Research Award. See Humbolt Award for details. In recognition of the new continuous flow instruments added to the UCSC stable isotope facility a Symposium is being run to highlight the new analytical capabilities. |